BD125X/2 / Bowie & Dick Test Pack Chemical Indicator designed for testing air removal efficiency of vacuum-assisted steam sterilizers. 100 % Lead Free.
Bowie-Dick test pack detects inadequate air removal and steam penetration in vacuum-assisted Steam sterilizers. Cycles of 16.9 min at 121 ยฐC, 4 minutes at 132 ยฐC and 3.5 minutes at 134 ยฐC.
Brand | Chemdye |
---|---|
Process | Steam |
Packaging | 20 |
Initial Color | Yellow |
Final Color | Black, Dark Brown |
Indicator Type | Type 2 Chemical Indicator |
Regulations | ISO 11140-1, ISO 11140-4 |
Field of activity | Dental, Healthcare, Pharmaceutical |
Characteristics
Type 2 Chemical Indicator.
Initial Color: yellow.
Final Color: dark brown/black (should change homogeneously in the
whole sheet).
Size: 16 mm x 125 mm x 125 mm (approximate).
Assembly: permeable barriers โ 1 silicone paper โ 1 test sheet โ permeable barriers. Wrapped with crepe paper.
Instructions for use
1. Place the Bowie-Dick Test Pack horizontally with its label facing up in an empty vacuum-assisted sterilizer chamber, in the bottom/front of the sterilizer rack, and over the drain.
2. Run the cycle for no longer than 3.5 min. at 134 ยฐC / 4 min. at 132 ยฐC / 16.9 min. at 121 ยฐC.