KBD8948X1 – Bowie-Dick Test Card Kit For Steam

KBD8948X1 / Bowie & Dick test card kit for Steam sterilization processes, includes Bowie & Dick-Test Card Chemical Indicators and holder.

Chemdye® Bowie & Dick Test Card has been designed to monitor the effectiveness of air removal in vacuum-assisted steam sterilizers operating at 134 ºC, 3.5 minutes / 132 °C, 4 minutes.

Brand Chemdye
Process Steam
Packaging 1 holder + 120 test cards
Initial Color Purple
Final Color Green
Indicator Type Type 2 Chemical Indicator
Regulations ISO 11140-1, ISO 11140-5
Field of activity Dental, Healthcare

Chemdye® Bowie & Dick Test Card should be used daily, after a sterilizer is installed or relocated, after malfunction, sterilization process failures and any major repairs of the sterilizer. Chemdye® Bowie & Dick Test Card consists of a Type 2, metals free chemical indicator. Cycle information can be recorded on the back of the indicator, thus allowing a permanent record of the process.

Chemical indicator changes from purple to green when processed. An heterogeneous color change, i.e the presence of purple color, indicates presence of an air pocket during the sterilization cycle, thus indicating sterilizer malfunction.

Chemdye® Bowie & Dick Test Card must be used daily, before the first sterilization cycle of the first load. A shortened cycle (i.e., a cycle omitting the post vacuum drying phase) should be run first to properly heat up the sterilizer.