Bowie Dick Test Card Holder, Steam Sterilisation

Bowie & Dick test card holder Steam sterilisation processes.

Chemdye® Bowie & Dick Test Card has been designed to monitor the effectiveness of air removal in vacuum-assisted steam sterilisers operating at 134ºC, 3.5 minutes / 132°C, 4 minutes.

Description

Bowie & Dick test card holder Steam sterilisation processes.

Chemdye® Bowie & Dick Test Card has been designed to monitor the effectiveness of air removal in vacuum-assisted steam sterilisers operating at 134ºC, 3.5 minutes / 132°C, 4 minutes.

Chemdye® Bowie & Dick Test Card should be used daily, after a sterilizer is installed or relocated, after malfunction, sterilisation process failures and any major repairs of the sterilizer. Chemdye® Bowie & Dick Test Card consists of a Type 2, metals free chemical indicator. Cycle information can be recorded on the back of the indicator, thus allowing a permanent record of the process.

Chemical indicator changes from purple to green when processed. An heterogeneous color change, i.e the presence of purple color, indicates presence of an air pocket during the sterilisation cycle, thus indicating steriliser malfunction.

Chemdye® Bowie & Dick Test Card must be used daily, before the first sterilisation cycle of the first load. A shortened cycle (i.e., a cycle omitting the post vacuum drying phase) should be run first to properly heat up the steriliser.

Instructions for use

1. Insert the Chemdye® Bowie & Dick Test Card in a Chemdye® holder with the chemical indicator side facing up. Place it in horizontal position on the bottom rack, near the drain of the empty steriliser.

2. Run the cycle for no longer than 3.5 minutes – 134ºC / 4 min. – 132 ºC. Long drying phases may cause the indicator to turn to other uniform shades of green.

3. After the cycle is over, remove and examine the chemical indicator.

4. Introduce data indicated at the back of the card and keep it as a permanent record.